Group 6: Radiation
|
|
40xxx
|
|
|
|
|
ProduktName
|

Zoom | Type-No.: 6017.0000 PAR Sensor The sensor measures radiation within the photosynthetic relevant spectrum. The sensitivity corresponds the optimum efficiency of chlorophyll. The measuring results enable a reliable assessment of development conditions of plants.
| Meas. range: |
0...approx. 250 W/m2
|
| Spectral sensitivity: |
380...700 nm
|
| Sensitivity-max. at: |
420 nm and 600 nm
|
Operating temp.:
|
-20°C...+60°C
|
Output:
|
0...5 V or 0/4...20 mA
|
Power supply:
|
+9...+30 V DC
|
Diffusor:
|
PTFE
|
Dome:
|
PMMA (UV- pervious)
|
Cosine-correction:
|
error f2 < 3%
|
Linearity:
|
< 1 %
|
Absolute error:
|
< 12 %
|
Offset (E=0):
|
< 20 mV
|
Dimensions:
|
∅ 42 x 70 mm
|
Weight:
|
approx. 0.10 kg
|
|

Zoom | Type-No.: 6018.0000 PAR Sensor as type 6017, but with enhanced optical features suitable for demanding scientific requirements.
6018.1000
|
Options: with mounting plate
|
6018.2000
|
with ventilation device
|
Meas. range global: |
0...approx. 250 W/m2
|
Spectral sensitivity: |
380...700 nm
|
Sensitivity-max. at:
|
420 nm and 600 nm
|
| Operating temp.: |
-20...+60°C
|
| Output: |
0...5 V or 0/4...20 mA |
Power supply:
|
+9...+30 V DC
|
Diffusor:
|
PTFE
|
Dome:
|
optical glas
|
Cosine-correction:
|
error f2 < 3%
|
Linearity:
|
< 1 %
|
Absolute error:
|
< 10 %
|
| Offset (E=0): |
< 10 mV
|
Dimensions:
|
∅ 80 x 82 mm
|
Weight:
|
approx. 0.30 kg
|
|
|
techn. Daten
|
| |
|